Zeiss Xradia Crystal CT
Available

Zeiss Xradia Crystal CT

Zeiss Xradia Crystal CT

Request Quote
Commercial Rate

This instrument can be used in two different modes: Absorption Contrast Tomography (ACT): the sample is rotated in small steps around the vertical axis and absorption images are recorded for each orientation; those images are then recombined into a 3D reconstruction of the mass-density of the object. Diffraction Contrast Tomography (DCT): the diffracted beams generated by the crystalline structure of the polycrystalline sample are intercepted by a planar detector and used, along with the ACT signal, to reconstruct the 3D microstructure of the sample, including the crystallographic orientations of the individual grains. In the DCT mode, the standard data acquisition mode is augmented by three advanced acquisition modes: helical phyllotaxis for cylindrical samples with high aspect ratios; helical phyllotaxis raster, for samples that are wider than the field of view, and helical phyllotaxis HART, for flat or plate-like samples. All modes allow for grain microstructures down to about 20 μm grain size to be studied in a non-destructive way, opening the path to in-situ observations.

Capabilities

  • Absorption Contrast Tomography (ACT)
  • Diffraction Contrast Tomography (DCT)
  • helical phyllotaxis
  • helical phyllotaxis raster
  • helical phyllotaxis HART

Specifications

ACT resolutionSpatial resolution: 0.95 μm; achievable reconstruction voxel size at working distance: 0.5 μm at 0.5 mm; 0.8 μm at 2.5 mm; 2.5 μm at 12.5 mm; 4.0 μm at 25 mm; 12.1 μm at 100 mm
Detectorthe detector is a high speed large array CMOS flat panel detector with 3,072 x 1,944 pixels, providing a single field of view size of 144 mm (diameter) by 93 mm (height), with a maximum height of 165 mm.
Reconstructions ACTZeiss XMReconstructor 
Reconstructions DCTXNovo Technology GrainMapper3D (allcrystal symmetries available)
Acquisition SoftwareZeiss Scout-and-scan control system
X-ray Tubethe x-ray source is a spot stabilized, sealed transmission x-ray tube with a voltage range of 30-160 kV and a maximum output of 10 W; set of 13 filters to tune the beam based on sample size and density.
Sample stageup to 25 kg sample mass capacity
System stabilitygranite base vibrational isolation; thermal environment stabilization; low noise detector
DCT resolutiongrain angular resolution 0.1°; grain detectability 20 μm;
VisualizationsORS Dragonfly Pro for 3D visualization and analysis

Similar Instruments

Zeiss Axio Scan.Z1 Slide Scanner
available
Imaging

Zeiss Axio Scan.Z1 Slide Scanner

ZeissAxio Scan.Z1 Slide Scanner

$0.00/hr
Harvard University
0.0(0)
Zeiss Observer Z1
available
Imaging

Zeiss Observer Z1

ZeissObserver Z1

$75.00/hr
Rutgers University
0.0(0)
Zeiss Axiovert 200M
available
Imaging

Zeiss Axiovert 200M

ZeissAxiovert 200M

$75.00/hr
Rutgers University
0.0(0)
Zeiss LSM 510 META (Inverted)
available
Imaging

Zeiss LSM 510 META (Inverted)

ZeissLSM 510 META

$75.00/hr
Rutgers University
0.0(0)