Available
Gel Permeation Chromatography (GPC)
Waters Alliance 2695 (2025)
Request Quote
Commercial Rate
Gel Permeation Chromatography (GPC) also referred to as Size Exclusion Chromatography (SEC). THF ONLY BASED SYSTEM. Located in Olney G-12. Performed on a Waters Alliance 2695 HPLC system with PDA detector and refractive index detector. Uses Waters Styragel HR columns 1,3, and 4 for molecular weight determination of up to 600K. System uses 100% tetrahydrofuran (THF only) as the mobile phase at 0.9ml/min flow rate. Each sample must be dissolved in 100% tetrahydrofuran for 24 hours before analysis at 10 to 20 mg/ml, and filtered through a 0.45 micron nylon or PTFE filter. 105 sample carousel for automatic injections. Training takes 1-2 hours.
Capabilities
- ✓Gel permeation chromatography
- ✓Size exclusion chromatography
- ✓Molecular weight determination
- ✓Polymer analysis
- ✓Up to 600K molecular weight
- ✓PDA detector
- ✓Refractive index detector
- ✓105 sample carousel
Specifications
| manufacturer | Waters |
| model | Alliance 2695 |
| mobile phase | 100% tetrahydrofuran (THF only) |
| flow rate | 0.9 ml/min |
| columns | Waters Styragel HR 1,3,4 |
| molecular weight range | Up to 600K |
| sample prep | 10-20 mg/ml in THF, 24h dissolution |
| filter | 0.45 micron nylon or PTFE |
| carousel | 105 samples |
| software | Empower 2 |