The Bruker D8 Venture at UMass Dartmouth represents the cutting edge of single-crystal X-ray diffraction technology. This comprehensive guide covers everything from basic operation to advanced techniques for obtaining publication-quality structural data.
Equipment Access Information
This Bruker D8 Venture is available for collaborative research through the Capital Equipment Network.
View Equipment Details & Book TimeSystem Overview
The D8 Venture is a dual-source diffractometer equipped with both Molybdenum (Mo Kα, 0.71073 Å) and Copper (Cu Kα, 1.54178 Å) X-ray sources. This flexibility allows researchers to optimize data collection for different sample types, unit cell sizes, and research objectives.
Key Specifications
- Detector: PHOTON III CPAD detector with 4096 × 4096 pixels
- Goniometer: Four-circle kappa geometry with 360° continuous φ and ω rotation
- Temperature Range: 100-400 K using nitrogen cryostream
- Crystal Size Range: 10 μm to 1 mm
- Resolution Capability: Up to 0.45 Å with Mo source, 0.70 Å with Cu source
- Software Suite: APEX4 for data collection, SAINT for integration, SADABS for absorption correction
Choosing Between Mo and Cu Sources
Source selection is one of the most important decisions in X-ray crystallography experiments. The choice impacts data quality, experiment duration, and the types of information you can extract.
Molybdenum (Mo Kα) Source
Best for: Routine structure determination, larger unit cells, absolute structure determination, charge density studies
Advantages:
- Shorter wavelength penetrates deeper into crystals
- Lower absorption coefficients for most organic compounds
- Better for anomalous scattering experiments (heavier atoms)
- Can access higher resolution (smaller d-spacing)
- Preferred for protein crystallography at in-house sources
Disadvantages:
- Weaker scattering from light atoms (H, C, N, O)
- Lower signal-to-noise for small molecules
- Longer exposure times required
Copper (Cu Kα) Source
Best for: Small molecule crystallography, weakly diffracting crystals, compounds containing only light atoms, quick structure determination
Advantages:
- 4× stronger scattering than Mo for light atoms
- Shorter data collection times (often 30-50% faster)
- Better peak-to-background ratios
- Ideal for organic molecules and coordination complexes
- Lower power consumption
Disadvantages:
- Higher absorption by most elements
- Cannot reach as high resolution
- Problematic for crystals containing heavy elements (Br, I, metals)
- Unit cells limited to ~25 Å due to resolution constraints
Standard Operating Procedure
1. Sample Preparation and Mounting
Crystal quality determines data quality. Select a crystal with:
- No visible cracks or twinning
- Regular shape (blocky preferred over needle-like)
- Size between 50-300 μm for most applications
- Uniform optical properties under polarized light
Mounting procedure:
- Select appropriate loop size (crystal should fit with minimal oil)
- Apply thin film of paratone oil to loop
- Transfer crystal to loop using microtools or oil transfer
- Minimize mother liquor around crystal
- Flash-freeze immediately if using cryo-protection
2. Initial Centering and Video Microscopy
Proper crystal centering is critical for accurate data collection:
- Mount crystal on goniometer head
- Start nitrogen flow and allow temperature to stabilize (5-10 min)
- Use video microscopy to center crystal in all three dimensions (X, Y, Z)
- Verify centering by rotating goniometer 180° and checking offset
- Fine-tune using optical centering routine in APEX4
3. Unit Cell Determination
The D8 Venture uses a "matrix" approach to quickly identify the unit cell:
- Collect matrix frames (typically 12-36 frames, 0.5° each)
- Index reflections using APEX4 indexing algorithm
- Refine unit cell parameters
- Assess crystal quality (mosaicity, resolution limit)
- Check for twinning, ice rings, or powder contamination
Quality indicators to examine:
- Mosaicity < 1.0° indicates good crystal quality
- Clear, sharp diffraction spots
- No ice rings at 3.67, 3.44, or 2.67 Å
- Diffraction visible beyond 0.80 Å resolution
4. Data Collection Strategy
APEX4 generates an optimized data collection strategy based on:
- Unit cell parameters and crystal system
- Desired resolution limit
- Estimated exposure time per frame
- Required completeness and redundancy
Typical collection parameters:
- φ or ω scan width: 0.5° (routine) to 0.3° (high resolution)
- Exposure time: 10-60 seconds per frame depending on crystal diffraction strength
- Resolution: 0.70-0.84 Å for routine work, 0.45-0.60 Å for high-resolution studies
- Redundancy: 4-8× for routine structures, 8-12× for charge density or weak anomalous signal
Data Processing and Structure Solution
Integration with SAINT
After data collection, process frames using SAINT:
- Load raw data files into SAINT
- Refine unit cell using all frames
- Integrate intensities for all reflections
- Apply Lorentz and polarization corrections
- Generate HKL file with I, σ(I), and indices for each reflection
Key metrics to evaluate:
- Rint < 5% for high-quality data
- Completeness > 99% for routine structures
- I/σ(I) > 10 for strong data
Absorption Correction with SADABS
Apply empirical absorption correction:
- Import HKL file from SAINT into SADABS
- Select multi-scan absorption correction method
- Optimize correction using redundant reflections
- Examine Rint before and after correction (should decrease)
- Export corrected HKL file for structure solution
Structure Solution and Refinement
Structure solution can be performed using various methods and software packages:
SHELXT (intrinsic phasing):
- Best for most organic and organometallic structures
- Fast and robust for data to 0.84 Å
- Automatically locates most heavy atoms
SHELXD (Patterson methods):
- Ideal for structures with heavy atoms
- Uses anomalous scattering for phasing
- Requires data collected with appropriate wavelength
Refinement in SHELXL:
- Load initial structure solution
- Refine atomic positions and isotropic displacement parameters
- Locate and add missing atoms using difference Fourier maps
- Switch to anisotropic displacement parameters for non-H atoms
- Add hydrogen atoms in calculated positions
- Apply appropriate constraints and restraints
- Refine to convergence (ΔF/σ < 0.001, shift/esd < 0.1)
Advanced Techniques
Low-Temperature Data Collection
Collecting data at 100 K provides several advantages:
- Reduced thermal motion (better defined electron density)
- Higher resolution limits
- Decreased radiation damage
- Stabilization of air-sensitive compounds
Absolute Structure Determination
For chiral compounds without heavy atoms, use Cu Kα source:
- Enhanced anomalous scattering from light atoms (O, N)
- Collect complete sphere of data (not hemisphere)
- Merge Bijvoet pairs separately during integration
- Use Flack parameter or Hooft parameter to assess enantiomer
Variable Temperature Studies
The D8 Venture supports temperature-dependent studies:
- Phase transitions
- Thermal expansion coefficients
- Order-disorder transitions
- Temperature-dependent twinning
Troubleshooting Common Issues
Weak or No Diffraction
- Check crystal is actually present and not dissolved
- Verify crystal is centered in beam
- Try longer exposure times
- Switch to Cu source for organic crystals
- Ensure crystal is not dehydrated or degraded
Ice Formation
- Powder rings at 3.67, 3.44, and 2.67 Å indicate ice
- Increase cryoprotectant concentration
- Flash-freeze more quickly
- Remount crystal with less mother liquor
High Mosaicity (> 2°)
- Check for crystal cracking during freezing
- Verify goniometer centering is accurate
- Screen multiple crystals to find better quality
- Optimize crystallization conditions
Split Reflections or Twinning
- Use CELL_NOW to identify twin law
- Integrate using appropriate twin matrix
- Refine twin fraction in SHELXL
- Consider collecting data from a different crystal
💡 Pro Tip: Community Knowledge Sharing
Join the discussion on our equipment community tab to share protocols, ask questions, and learn from other researchers using the D8 Venture.
Conclusion
The Bruker D8 Venture represents a powerful platform for single-crystal X-ray diffraction studies. By understanding source selection, optimizing data collection parameters, and following best practices for structure solution, researchers can obtain publication-quality crystallographic data efficiently.
Remember that successful crystallography requires practice and patience. Don't hesitate to collect test frames, optimize conditions, and seek advice from experienced users through the Capital Equipment Network community.
Ready to Get Started?
Book time on the Bruker D8 Venture or explore other X-ray crystallography equipment available through our network.
Francesco Piscani
francesco@capneteq.com
Contributing author at Capital Equipment Network